APK-QY8 Dissolution Autosampler
Allpack APK-QY8 dissolution autosampler is composed of Huanghai APK-8A intelligent drug dissolution apparatus and APK-8A automatic sampling system. It is an 8-channel high-performance drug dissolution test device upgraded and developed in strict accordance with the current “Chinese Pharmacopoeia” and “Guidelines for Mechanical Validation of Drug Dissolution Apparatus”, and fully complies with data integrity specifications.
APK-QY8 Dissolution Autosampler Features:
1.Built-in three-level user authority management, audit tracking, and test record management.
2.Test reports can be printed or output to USB through the device interface. Common sample information management for easy retrieval and use.
3.Common test plan management for easy retrieval and use.
4.High-precision real-time constant temperature control, independent temperature measurement for each dissolution cup, with dual over-temperature protection and temperature calibration functions. Internally equipped with scheduled heating, users can set the system to start heating at a specified time.
5.Manual and automatic synchronous dosing function.
6.Manual and automatic sampling dual mode.
7.Split agitator design for quick replacement.
8.Automatic centering of the dissolution cup and agitator.
9.Sealed anti-evaporation cover design effectively reduces medium volatilization.
10.More convenient water circulation pipeline design, professional medical-grade quick connector.
11.High-precision syringe pump is used.
12.The sampling needle enters the dissolution cup during sampling to avoid turbulence.
13.Samples are replenished isothermally and in equal amounts.
14.Secondary filtration can be performed.
15.The system automatically rinses the pipeline before sampling and automatically empties the pipeline after sampling.
16.IQ/OQ certification documents are provided free of charge.
APK-QY8 Dissolution Autosampler Technical Specification:
Product model |
APK-QY8 |
|
Dissolution apparatus parameters |
dissolution channel |
8 channels |
Stirring paddle swing range |
≤0.5mm |
|
Basket swing range |
≤1.0mm |
|
Deviation of rotary rod and dissolution vessel axis |
≤1.0mm |
|
Speed setting range |
10~300RPM |
|
Steady speed error |
≤±1% |
|
Temperature range |
Room temperature~45.0℃ |
|
Temperature resolution |
0.1°C |
|
Temperature control accuracy |
≤±0.3℃ |
|
Sampling system parameters |
sampling channel |
8 channels |
Sampling times |
≤15 times |
|
Sampling timing time |
≤999 hours, 59 minutes and 59 seconds |
|
Sampling volume, replenishment volume |
≤10ml |
|
Sampling accuracy |
≤±1%, sampling volume 10ml, average value of each channel |
|
Sampling interval |
1 minute for the first time, 2 to 5 minutes thereafter (depending on the sampling volume) |
|
sample collection bottle |
120 bits |
|
power supply |
AC220V±10% 50Hz |
Laboratory Instruments Related Products